声明:本说明书著作权归杭州睿登科技有限公司所有,未经允许任何单位或个人不得用于商业用途。
46
1.4.2.2 Main Page Output Voltage and Current Setting
Output voltage and current setting operation video
:
https://drive.google.com/drive/folders/1NZW7k3-Kyc5YgIh-bWn8gBGaY5xo7_Lm?us
p=sharing
Press
or
button can switch the current range between 6A and 12A.
When you change the current range, the output will be turned off
Press
button to set the output current value,
you can use encoder potentiometer to adjust the
output value directly. And you will not set the value
which exceeds the limit in this way, press
button to move the cursor. Of course you can use
keypad to type in the value, and press
to
confirm, and it will save the set value and set current
range, if you set a value exceeds the limit, it will
prompt like what shows in Figure 1. If you set the wrong value, you can press
encoder potentiometer to cancel.
Press
button to set the output voltage value, the operation way is similar
to output current setting.
Press
+
button /
+
button to set the over current
protection/ over voltage protection value. The operation method is similar to output
current setting. If you want to set the over current auto cut off function, your OCP
value should be higher than the Current setting value.
When the device is under constant voltage mode, it will show
, and it will
show
when under constant current mode; when the device works normally it
will show
at protection status indication, when the actual output current value
is higher than the over current protection value(OCP), the output will be cut off
automatically, and show
, when the actual output voltage value is higher than
the over voltage protection value(OVP), the output will be cut off automatically, and
show
, when the system temperature is higher than 80
℃
, the output will be cut
off automatically, and show
.
1.4.2.3
Data Group Quick Storage and Call out
Data group quick store and call out operation video
:
https://drive.google.com/drive/folders/139d28IHpZEgpNEC6pfBVruNF3b9FRVWW?usp=sharing
Press
+ keypad button 1-9, you can store the output voltage value, output
Figure 1
15: 19: 17
O-12A
60.000
V
12.000
A
720.00
W
0
M
CV
000.000Ah
A
12.200
V
62.000
pr ot ect set
A
12.100
V
60.000
V&a set
I NPUT: 68. OOV
Set Er r or