Caution!
Medical electrical devices are subject to special precautions with regard to electromagnetic compatibility (EMC).
Portable and mobile radio frequency communication devices may affect medical electrical equipment. The ME device is intended for use in
an electromagnetic environment and for professional facilities such as industrial areas and hospitals.
The user of the device should ensure that it is used within such an environment.
Warning!
The ME device may not be stacked, arranged or used directly next to or with other devices. If the ME device must be operated near or sta-
cked with other devices, the ME device and the other ME devices must be monitored to ensure proper operation in this arrangement. This
ME device is only intended for use by medical professionals. This ME device is intended for use in professional healthcare facilities. This
device may cause radio interference or impair the operation of nearby devices. Appropriate corrective action may need to be taken, such as
redirecting or rearranging the ME device or shield.
The rated ME device does not exhibit any basic performance features in the sense of IEC 60601-1 that would present an unacceptable risk
to patients, operators or third parties should the power supply fail or malfunction.
Warning!
Portable RF communications equipment (radios), including accessories such as antenna cables and external antennas, should not be used
in closer proximity than 30 cm (12 inches) to parts and cables of the device specified by the manufacturer. Failure to comply may result in
reduced device performance.
Guidance and manufacturer‘s declaration - electromagnetic emissions.
The chargers are intended for use in the electromagnetic environment specified below. The customer or user of the device should ensure
that it is used in such an environment.
Guidelines and manufacturer‘s declaration – electromagnetic emissions
The ri-focus LED headlight and clar N head mirrors are intended for use in the electromagnetic environment specified below. The customer or user of the headlight
should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
HF emissions CISPR 11
Group 1
RF energy of the headlights is for internal use only. Thus, the RF emissions are very low and
are not likely to cause interference near other electronic devices.
HF emissions CISPR 11
Class B
The headlights are intended for use in all establishments, including residential areas and
those directly connected to a public supply network that also supplies buildings used for
residential purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Not applicable
Summary of Contents for ri-focus LED
Page 73: ...73 B MD II 2 REF C F 2002 96 EC 1 2...
Page 74: ...74 1 3 IEC 60601 1 2 1 4 Riester 1 4 1 6 6 1 4 2 Riester 1 4 3...
Page 75: ...75 1 4 4 1 4 5 1 4 6 1 5 Riester...
Page 77: ...77 2 3 2 3 1 2 3 2 ri focus 1 1 90 clar N 55 2 3 3 ri focus clar N AAA CR 123A 1 1 1...
Page 82: ...82 8 1 IEC 60601 1 30 12...
Page 83: ...83 ri focus Clar N CISPR 11 1 CISPR 11 B IEC 61000 3 2 IEC 61000 3 3...
Page 87: ...87 8 1 9 2 B RIESTER Rudolf Riester GmbH Repairs dept RR Bruckstr 31 D 72471 Jungingen Germany...