34 -
|
Page Rigel Medical
62353+
user Manual V1.4
Warning
Mains voltage applied to appliance.
This example uses the Semi Automatic Mode and the user has the opportunity to
determine the moment at which the leakage measurements start. In fully automatic
mode , the Rigel 62353 Plus will start the measurements immediately after the EUT has
been switched on.
The initial powered test is a load test. The Rigel 62353 Plus will check for a short circuit
EUT and ensure the applied load (EUT) is less than 16 A.
In case of a short circuit in the EUT, the Rigel 62353 Plus will not power-up and a
warning message is displayed.
Following the Load Test, the individual leakage tests are performed, measured and
displayed.
The applicable SFC is displayed in the bottom left side of the display. The following is a
typical sequence for a Class I device.