50 -
|
Page Rigel Medical
62353+
user Manual V1.4
button (F3)
Press
to set the Test Duration, Insulation test voltage (500 / 250VDC) and Pass/Fail limit. Use
the up, down, left and right keys to highlight and set. Press
button (F4) when
complete. Pressing
button (F3) will exit Settings mode without saving changes.
For both Class I and Class II appliances, connect ALL Patient Connections or Applied
Parts to the Applied Part Module and fit the Appliance mains plug into the EUT outlet.
All Patient Connections and Applied Parts are grouped together which
eliminates the requirement to configure the Applied Parts as the Rigel
62353 Plus will short together all connections as part of this test. For
this reason, the Applied Part configuration key is not available.
For Class I equipment use the earth bond lead (from the black socket) and connect to a
conductive part of the enclosure. For measurements on conductive non-earthed parts,
the test needs to be repeated using the same probe.
For Class II devices use the earth bond lead (from the black socket) and connect to the
enclosure preferably encapsulated within conductive foil.
Press the green START key to begin the test. The test will run for the duration set or
until the red STOP key is pressed.