Rigel SafeTest 60 Instruction Manual
–
Revision 1.2
12
The SafeTest 60 will automatically start the test when it is selected and will
automatically stop by pressing the LEAKAGE, INSULATION, POINT TO POINT
or HOME icon.
Connection between EUT and SafeTest 60;
EARTHBOND TESTING
Each time the earthbond probe is placed on a new test point, the
zap circuit will be reactivated, ensuring accurate readings at
every measurement point.
Do not exceed the maximum permitted voltage of 30 V AC/DC with
respect to earth potential! Electric Shock danger!
To compensate for the test cable resistance, connect the test cable between the
black earthbond socket and the EUT earth, then press the
button on the
front panel.
When the lead compensation is activated, the
icon will appear on the
screen.
To remove the lead compensation, remove the probe and press the
button.
When different earthbond cables are used, the NULL function must
be repeated for each different cable.
Switching off the Rigel SafeTest 60 will not
cancel the ‘probe zero’.