Rigel SafeTest 60 Instruction Manual
–
Revision 1.2
14
To activate single fault conditions, use the
and
buttons on the front
panel.
opens the EARTH fault condition relay whilst the
opens the
NEUTRAL fault condition relay.
The leakage screen will indicate the current state of the selected fault conditions;
Connection between EUT and SafeTest 60;
ENCLOSURE LEAKAGE
ALL Applied Parts
Leakage Current Type
NC
SFC
Earth Leakage (3rd edition)*
5mA
10mA
Earth Leakage (General)
0.5mA
1mA
Enclosure Leakage
0.1mA
0.5mA
* The pass fail limit for Earth Leakage in the 3rd edition of IEC 60601 has been increased from
500μA under normal condition to 5000μA for class I equipment with NO exposed metal parts that
may become live when a fault appears.