Rigel SafeTest 60 Instruction Manual
–
Revision 1.2
28
10
Technical Specifications
Earth Continuity
Pre-pulse
65-25A peak current,
(0.1 to 0.8 Ω respectively)
Pulse shape
exponential decay
Decay time
200
– 550µs to 5% of peak current , (0.1 to 0.8 Ω respectively)
Method
2 wire,
Measurement Current
>± 200mADC into 2
Ω
Max Test Voltage
4-24Vrms o/c
Measuring Range (low range)
0.001
– 0.999
Ω
Resolution
0.001
Ω
Measuring Range (mid range)
1.00
– 9.99
Ω
Resolution
0.01
Ω
Measuring Range (high range)
10.0
– 19.9
Ω
Resolution
0.1
Ω
Accuracy
± 1% of value, ± 5m
Ω
Circuit Protection
Test inhibited if
≥ 30VAC or DC at 4mm inputs
Insulation Resistance
Measurement
EUT to Earth
Voltage
500VDC @1mA.
Maximum O/C Voltage
<600VDC
Range
100K - 20M
Ω
± 5% ± 2 digits
Resolution
0.01M
Ω
Short circuit current
<2mA
Powered Leakage Measurements
IEC 62353
Equipment Leakage (Direct)
IEC / AAMI 60601
Earth + Enclosure Leakage
NFPA-99
Chassis Leakage
IEC 61010
Touch Leakage, Touch Voltage
Test Voltage
Mains Supply Voltage
Measuring Range
0.1-
9999μA (0.1-8000μA typical for IEC61010)
Measurement/Display Resolution
0.1μA
Accuracy
± 2%, ± 5μA
Mains Reversal
Soft key
Single Fault Conditions
Open neutral and Open earth via soft key
Frequency Response
IEC 60601
– 62353, NFPA-99, and IEC 61010 selectable
Voltage measurement
Application
L-N, L-E, N-E and touch voltage (IEC 61010)
Range
0.0V
– 300VAC
Resolution
0.1V
Accuracy
± 2% ± 2 digits (between 10V
– 270VAC)
Mains frequency
45.0
– 66.0Hz
Resolution
0.1Hz
Accuracy
Unspecified
EUT Load Current Measurement
Range
0.0A
– 20.0A
Resolution
0.1A
Accuracy
±5% ± 2 digits
Power Source
Maximum current rating
20A @ 120V / 16A @ 230V
Duty cycle (@21°C ambient)
16A to 20A, 3 min. on/ 10 min. off
10A to 15A, 3 min. on/ 5 min. off
0A to 10A, continuous
Mains power
90-264V 48-64Hz
Weight
1.1kg / 2.5lbs (unit)
2.2kg / 5lbs (complete with accessories)