7 |
P a g e
R i g e l M e d i c a l U n i - T h e r m u s e r M a n u a l V 1 . 0
1
Getting started
The Rigel Uni-Therm has been designed to ensure that the user is able to test: quickly,
easily and safely. The interfaces to the Rigel Uni-Therm have been segregated into a
high-voltage area (side panel) and a safe low voltage area (front panel). This
arrangement also ensures that test leads carrying high frequency power are kept as
short as possible, and directly facing the equipment under
test.
1.1
Connecting your Analyser
Elevate the Rigel Uni-Therm on its two hinged feet
at the front base.
Ensure none of the large base and rear ventilation
inlets are obstructed during operation,
1.2
Rear panel
Connect the mains cable to the IEC connector on the rear
panel.
1.3
Front connection panel
For automatic control during HF leakage and Power tests, connect the ESU footswitch
COAG and CUT contacts to the blue and yellow terminals.
It is strongly advised to always use the Rigel Uni-Therm’s built-in, automatic control
rather than manually switching the output of the electrosurgical diathermy machine
under test.
For REM / plate security test use the black terminals.