R i g e l M e d i c a l U n i - T h e r m u s e r M a n u a l V 1 . 0
2
Manual Mode
The manual mode provides the user with the features for testing a specific individual
function and or test condition; for example to aid fault diagnostic procedures. These tests
are available from the home screen.
Available tests are:
REM test
HF Leakage
Power Test
2.1
Plate Security (REM) test
This test will control a motor driven potentiometer to trigger the neutral (patient plate)
alarm on electrosurgical generators by simulating fault conditions (ie resistance too high
or too low, a variation in resistance etc). The variable resistance (0 – 475 Ω) is
connected to the two black connectors on the front panel.
Select the REM test from the home screen shown below.
During the initial REM test selection, the Rigel Uni-Therm will auto-calibrate the REM
potentiometer;