Once the calibration has completed, the REM test screen will appear;
Select automatic (up or down) or manual (up or down) control by using the dedicated
AUTO MAN key. This key is also used to set the limits. Once the SET LIMITS field is
highlighted in green, use and press the rotary encoder to manipulate the UP and DOWN
limits.
Press the Green button to start the test (for automatic mode only). Then capture the
alarm value by pressing the dedicated ‘Capture Alarm’ key on the screen.
In MANUAL UP and DOWN mode, the rotary encoder is used to directly control the
potentiometer. The alarm value is then captured by pressing the dedicated ‘Capture
Alarm’ key on the screen.
2.2
HF Leakage
The HF Leakage test measures the HF leakage current in various test configurations
(see
), and compares the result to a user set pass/fail value.
Select the HF LEAKAGE test from the main menu, shown below.
Use the rotary encoder to navigate the screen