or the IEC 60601-2-2 reference details (SHOW DETAIL). The rotary encoder can be
used to scroll the different test setups, or to SET LOAD and LEAKAGE LIMIT values.
The selected field is highlighted in red. Press the rotary encoder to activate the field (field
turn blue) and use the rotary encoder to change the settings. Press the rotary encoder
again to confirm
Return to the DIAGRAM mode by using the SHOW DIAGRAM fast key.
Start the test by pressing the RUN button, and confirm with the green START button on
the front of the Uni-Therm;
The Rigel Uni-Therm could display the following details after the test;
Press the HOLD key to freeze the reading on the screen.
The test can be aborted by pressing the red STOP button on the front of the Uni-Therm.
After a test is completed (either PASS or FAIL) press the green START button on the
front of the Uni-Therm.