18 |
P a g e
R i g e l M e d i c a l U n i - T h e r m u s e r M a n u a l V 1 . 0
3 Automatic Mode
The automatic mode allows the user to select a test sequence which will carry out a pre-
determined set of tests against user definable limits. Test results are stored at the end of
the test for future reference. To view or transfer stored data, please refer to chapter 4)
To enter the automatic mode, select the MENU key (F1) from the home screen and then
AUTOMODE;
In the following screen, an asset ID number can be entered using the rotary encoder,
barcode scanner or a PS2 keyboard.
When using the barcode scanner, ensure the barcode scanner is paired with the Rigel
Uni-Therm, see
. (The Rigel Uni-Therm will only support the Seaward barcode
scanner, p/n 339A923). When the barcode is ready to scan, the word “OK” will appear in
the asset ID field.
Enter data with the rotary encoder (push to select character) or keyboard (note that the
enter key is not required). Press NEXT to choose an available test sequence;