Error Messages and Trouble Shooting
59
58
Early applying blood sample or control solution
Strip Error - Er1
Meter Error - Er2
Signal Error - Er3
The blood sample or control solution are applied before the meter is ready. Repeat
the test with a new test strip. Add sample after the " " symbol appears.
Test strip inserted has been used or damaged. Please use a new test strip from vial.
Meter has malfunctioned. Do the quality control test or reinstall the batteries to see if
the meter works properly.
If error screen still appears, contact Bionime Customer Service.
When you do the test, the meter find the signal is unreasonable. Please do a new
test again with a new test strip.
Limitations
- Store the strips in the original capped vial at temperatures between 4
℃
to 30
℃
( 39
℉
to 86
℉
)
and relative humidity below 90 %. Do not freeze.
- Blood Glucose Test strips GS300 are designed for using with capillary whole blood
samples. Do not use serum or plasma samples.
- Inaccurate test results may be obtained at high altitude more than about 3,048 meters (10,000 feet )
above sea level.
- Hematocrit should be between 30 %~55 %. If you do not know your hematocrit, ask your healthcare
professional.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- Blood Glucose Monitoring System has not been validated for use on neonates.
- The glucose test may be interfered under abnormal concentration of
Ascorbic acid
≧
5 mg/dL ( 0.28 mmol/L),
Dopamine HCl
≧
2 mg/dL ( 0.11 mmol/L),
L-Dopa
≧
3 mg/dL ( 0.15 mmol/L),
Glutathione reduced
≧
70 mg/dL ( 2.28 mmol/L),
Hemoglobin
≧
6,000 mg/dL( 0.94 mmol/L),
Uric Acid
≧
10 mg/dL( 0.59 mmol/L)