41
40
Recalling Test Result and Average
37
Recalling Test Result and Average
3. Non-Averaging set and canceled:
You may select the unwanted or doubtful test results, to exclude them from the average calculation, after your every test. After you finished
your test in the test mode, you can exclude the unwanted result. Keep pressing the Right butten until you see the symbol “NO AVG” in the
top right cornetr of the screen. By keeping the Right button pressed and pressing the main button simultaneously to confirm the value as
“Non average”.
Now this value is flagged by the symbol “NO AVG” in the up per right corner of the screen.
If you want to cancel this non-averaging setting, you have to stay in the test mode. First keep pressing the Right button until the “NO AVG”
symbol changes to “AVG” in the up per right corner of the screen. To confirm this original status you again simultaneously press the Main
button to confirm the “AVG” status. Now the value will be stored as a regular value and included in the average values.
4. Quick Searching: If you want to see all values automatically displayed in sequence, first enter the memory mode. Then
keep pressing the Right or Left button for 2 seconds. (Right button is for searching the tests the latest value to the
oldest one; by using the Left button you can scroll from the oldest to the latest one). Anytime you want to stop just
release the pressed button. It will stop and display the current displayed value. You can use this way to find a particular
reading you want to check.
5. Reading right after test: If you just finish the test, press the Main button to enter the MEM (memory) screen and review
the latest test result.
NOTE
Your Control Solution results will be automatically non-averaged without selecting, but still can be recalled. The
Control Solution Test results will show with “CS” symbol under memory screen.
PRECAUTION
-
You have to set the time and date to activate the average function.
- The " " , " " results, the Control Solution results and the test result made out of normal temperature range
NOTE
- The average function is related to the time setting. You must set the time right and have enough time intervals on the basis of current
inquiry time, to make sure that the average test results will show. Use the 14-day average for example; if your current inquiry time is
2010/1/30, then that you must check that you have tested blood glucose in the past 14 days before 1/30, including today. If not, the 14-
day average will show no figure.
- The Non-averaging/re-averaging function only works in the test mode and when you get the value. In the MEM mode, you can not
perform this action.