Understanding Test Results and Messages
About Quality Control Testing
RIGHTEST Blood Glucose Meter displays results between 10 and
600 mg/dL (0.6 and 33.3 mmol/L).
If the test result is below 10 mg/dL (0.6 mmol/L) , " " will appear on the
screen. Please repeat your test with by a new test strip.
If you still get a " " result, contact your healthcare professional.
If the test result is above 600 mg/dL (33.3 mmol/L), " " will appear on the
screen. Please repeat your test again with a new test strip.
If you still get a " " result, contact your healthcare professional.
What is a Quality Control Test ?
To ensure proper monitoring function, it is necessary to regularly perform a quality control test.
Use one of the RIGHTEST Control Solutions when testing your RIGHTEST Blood Glucose
Monitoring System in the Control Solution Mode. If the test result is within the Control Solution
Range printed on the strip vial label, RIGHTEST Blood Glucose Monitoring System passes the
Quality Control Test and your RIGHTEST Blood Glucose Monitoring System is working properly.
Control Solution Range:
Example of Control Solution Range printed on your test strip vial label.
NOTE
- If you want to purchase new control solutions, please contact your local Bionime distributor
or email to rightest@bionime.com .
Level 1
Level 2
Level 4
Control Solution Range mg/dL
Normal (Level2): 88 - 120 mg /dL
Hing (Level4): 250 - 340 mg/dL
mmol/L
4.9 - 6.7 mmol/L
13.9 - 18.9 mmol/L