Performing a Quality Control Test
Performing a Quality Control Test
1. Take one Test Strip from the vial and close the vial cap immediately.
2. Insert the Test Strip horizontally (not vertically) with the colored view
window facing up into the test strip port.
3. While the blood drop icon is flashing on the display window, press
and hold the Main button for at least 3 seconds until the " " and
" L2 " symbol appears.
4. Keep pressing the main button for another 3 seconds to select the
" " level you want test (Level 2-Normal, Level 4-High) Ensure that
the control solution level on the control solution bottle matches the
level you are testing.
5. You will see the blinking " " icon and " " icon on the screen,
prompting you to apply the corresponding level of RIGHTEST Control
Solution.
6. Shake the bottle of RIGHTEST Control Solution well before opening
the cap. Place the cap on a flat surface.
7. Drip a drop of control solution onto the top of the cap.
8. Gently touch the sample port of the strip with the control solution from
the top of the cap.
9. The screen will display the count time starting from 5 (you will hear a
beep if the volume is turned on).
1
2
9
OPEN
6
7
3
10. Tightly replace the cap on the Control Solution bottle.
11. The control solution result will appear. Compare your Quality Control Test result
to the Control Solution Range printed on RIGHTEST Blood Glucose Test Strip
vial label.
CLOSE
10
CAUTION
- Your Control Solution Test results will not be included in the average calculations, however,
they can still can be recalled and viewed. The Control Solution Test result will be shown
with the " " icon on the screen.
- The Control Solution Test should be conducted between 6 - 44°C (43 - 111°F).
- Before " " and " " appear, do not touch the control solution to the sample port on the
strip. RIGHTEST Blood Glucose Meter is performing an internal check. Touching the control
solution to the sample port before prompted will result in an error message:
" " and " " and be accompanied by beeps (if volume is turned on).
- Do not drip the control solution on to the sample port of the test strip directly. The reagent
on the strip might leak into the bottle of control solution and may cause the degeneration
of the control solution. This could contaminate the meter via the test strip port.
- Keep the test strip port clean and dry. Clean immediately if the test strip port is
stained or is overly exposed to moisture.
- Do not touch the tip of the control solution bottle. If the tip is touched, clean
with water.
8