NOTE
- Clean and disinfect the outside of the device only. Do not remove battery cover when
cleaning and disinfecting.
To clean the meter:
1. Throughly wipe the entire surface of the meter with disinfecting
wipes listed above to clean any possible dirt, dust, blood and other
body fluids.
To disinfect the meter:
2. Take another disinfecting wipe and wipe the meter thoroughly.
(Note: All blood and body fluids should be cleaned from surface
before performing the disinfecting procedure)
3. Allow the surface to remain wet for 2 minute.
4. Allow to air dry.
NOTE
- Your RIGHTEST Blood Glucose Meter has been tested to ensure that there is no change
in the performance or external materials of the device after 10,000 cleaning cycles and
10,000 disinfecting cycles. The testing simulates high frequency of cleaning and
disinfecting cycles around 5 years..
2
1
3
Allow to remain
wet for 2 minute
1
2
3
Allow to remain
wet for 2 minute
To clean the lancing device:
1. Throughly wipe the entire surface of the lancing device with
disinfecting wipes listed above to clean any possible dirt, dust, blood
and other body fluids.
To disinfect the lancing device:
2. Take another disinfecting wipe and wipe the Lancing device
thoroughly. (Note: All blood and body fluids should be cleaned from
surface before performing disinfecting procedure)
3. Allow the surface to remain wet for 2 minute.
NOTE
- Your RIGHTEST Lancing Device has been tested to ensure that there is no change in the
performance of the device after 550 cleaning cycles and 550 disinfecting cycles. The
testing simulates 2 cleaning and disinfecting cycles per week over the typical life of the
meter (5 years).
Maintain the Products
Maintain the Products