53
52
Error Messages and Troubleshooting
Error Messages and Troubleshooting
Strip Error
When a test strip is not inserted correctly, "Strip Error" will flash on the screen.
Please reinsert the strip following the instructions on page 27 (Performing a
Blood Glucose Test). If after following the instructions, "Strip Error" appears
again, you may be using the wrong test strip. Please check to make sure you are
using a RIGHTEST Test Strip.
Sampling Error
Please do not apply the blood to the sample port of the test strip before the
meter displays " ". Discard the test strip If the meter shows " ".
Battery Error
1. The " " and " " symbols appear and the meter will not function when
the battery is low. Change the batteries immediately.
NOTE
- The backlight function operates off of its own battery source. If the
backlight is no longer functioning, replace the backlight battery (Refer to
page 11 to find the location of the backlight battery)
- When the backlight no longer operates, the meter will still function for
testing purposes until the " " and " " symbols appear as
described above.
Er1
-
The inserted test strip has been used or damaged. Please use a
new test strip.
Er2
-
The meter has malfunctioned. Do a Quality Control Test or reinstall
the batteries to check if the meter works properly.
Er3
-
The signal transmission is disrupted. Repeat the test using a new
test strip.
Er4
-
The blood sample volume is insufficient. Repeat the test using a
new test strip.
Er5
-
An issue calibrting the Meter has occurred.Please follow the steps
below:
1. Remove the test strip from the meter.
2. Turn off the meter.
3. Press the mailn button to turn ON the meter.
If after turning on the Meter, you do not see Er5, your meter is
functioning properly and able to perform a test.
Meter Malfunction
If the RIGHTEST Blood Glucose Meter will not turn on, please follow the steps below:
1. Open the battery cover and remove the batteries.
2. Wait for 5 minutes and reinsert the batteries as described as instructed on page 16, (Meter
Activation and Changing the Battery).
The meter should work normally after finishing the above steps. If the meter still does not work,
please contact your local local Bionime distributor or email to rightest@bionime.com .
Summary of Contents for GT200
Page 2: ......