Performing a Test

 

 

5

1

2

1) Wash and dry your hands. Take one test strip from the vial. Close the vial cap immediately.
2) Insert the test strip into the strip port on the meter with the view window facing up. Push the test strip in until you hear 

a click. The meter will automatically detect the code number.

3) When you see the flashing blood drop icon on the display window, hold the lancing device to your fingertip and press 

the release button. 

4) Sample Size Example

4) A blood drop size between 0.75 

µL

 and 3.0 

µL

 is required for testing. See Sample Size Example Chart for drop sizes. 

Blood drop sizes greater than 3.0 

µL

 may contaminate the meter.  A minimum of a 0.75 

µL

 drop is required for 

accurate results. 

        0.75 

µL

       1.0 

µL

        1.5 

µL

        2.0 

µL

        3.0 

µL

  5) Touch and hold the drop to the edge of sample port until you hear a " beep " and the view window is filled with blood. 

If the view window is not completely filled with blood, the test will not start. Discard the test strip and repeat the test 
with a new test strip.

  6) The countdown will begin on the display window. After 5 seconds, your test result will appear. For more information 
      about your test results, please see your User’s Manual.
  7) Rotate the test strip counterclockwise and pull up simultaneously to remove the test strip from the meter. Please 

follow your local regulations to properly discard the used test strip.

  8) Pull off the adjustable depth cap. Without touching the used disposable lancet, stick the lancet tip into the protective 
      cover.
  9) Hold the release button in one hand and pull the plunger with the other hand. This will safely eject the used 

disposable lancet.

10) Discard the used disposable lancet into an appropriate puncture-proof or biohazard container.
11) Replace the adjustable depth cap after finishing the test.

5a

7

6

5b

Alternative site testing: palm or forearm blood sampling 

1. Choose the clear cap and follow step 1 to 5 of the above section " 

Performing a Test ". 

2. Massage the puncture area of your palm or forearm for a few 

seconds. 

3. Immediately after massaging the puncture area, press and hold 

the lancing device with the clear cap against your palm or forearm. 

4. Press the release button.
5. Continue holding the lancing device against your palm or forearm. 

Gradually increase pressure for a few seconds until the blood 
sample size is sufficient. 

6. Follow step 9 of the above section to discard the used disposable 

lancet.

NOTE: DO NOT test on the palm or forearm if you are testing for  
            hypoglycemia (low blood glucose).

3a

3b

8

Releasing 

button

Pressing for

a few seconds

For more information on how to use your meter and understanding your test results, please see your GT333 Blood 
Glucose Monitoring System User's Manual.

4

5

9

10

Rev. Date:2019-10

101-3GM236-

XXX

EN

BIONIME CORPORATION

No. 100, Sec. 2, Daqing St., South Dist., 
Taichung City 40242, Taiwan (R.O.C)
Tel: +886 4 23692388
Fax: +886 4 22617586
E-mail: info@bionime.com
http://www.bionime.com

Bionime GmbH
Tramstrasse 16, 9442 Berneck, Switzerland
E-mail: info@bionime.ch

0197

Rotate

Palm

Forearm

Reviews: