RIGOL
2-76
DG900 Programming Guide
Parameter
Name
Type
Range
Default
[<n>]
Discrete
1|2
1
<seconds>
Real
10 ns to (0.625×pulse width)
20 ns
Remarks
The rise time is defined as the time required for the pulse amplitude to rise from 10% to 90%; the fall
time is defined as the time required for the pulse amplitude to fall from 90% to 10%.
When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related parameters of CH1 by default.
The ranges of the rise time and fall time are limited by the current waveform frequency and pulse
width. When the specified value exceeds the limits, the instrument will adjust the edge time
automatically to make it match the specified pulse width.
Example
:SOUR1:FUNC:PULS:TRAN 0.000000035 /*Sets the pulse rise time and fall time of CH1 to 35 ns.*/
[:SOURce[<n>]]:FUNCtion:PULSe:TRANsition:LEADing
Syntax
[:SOURce[<n>]]:FUNCtion:PULSe:TRANsition:LEADing {<seconds>|MINimum|MAXimum}
[:SOURce[<n>]]:FUNCtion:PULSe:TRANsition:LEADing? [MINimum|MAXimum]
Description
Sets the pulse rise time of the specified channel.
Queries the pulse rise time of the specified channel.
Parameter
Name
Type
Range
Default
[<n>]
Discrete
1|2
1
<seconds>
Real
10 ns to (0.625×pulse width)
20 ns
Remarks
The rise time is defined as the time required for the pulse amplitude to rise from 10% to 90%.
When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related parameters of CH1 by default.
The range of the rise time is limited by the current waveform frequency and pulse width. When the
specified value exceeds the limits, DG900 will adjust the edge time automatically to make it match the
specified pulse width.
Return Format
The query returns the pulse rise time in scientific notation, with 7 effective digits. For example,
3.500000E-08 (the pulse rise time is 35 ns).
Example
:SOUR1:FUNC:PULS:TRAN:LEAD 0.000000035 /*Sets the pulse rise time of CH1 to 35 ns.*/
:SOUR1:FUNC:PULS:TRAN:LEAD? /*Queries the pulse rise time of CH1 and the query returns
3.500000E-08.*/
[:SOURce[<n>]]:FUNCtion:PULSe:TRANsition:TRAiling
Syntax
[:SOURce[<n>]]:FUNCtion:PULSe:TRANsition:TRAiling {<seconds>|MINimum|MAXimum}
[:SOURce[<n>]]:FUNCtion:PULSe:TRANsition:TRAiling? [MINimum|MAXimum]
Description
Sets the pulse fall time of the specified channel.
Summary of Contents for DG952
Page 2: ......
Page 168: ......
Page 181: ...Chapter 4 Programming Examples RIGOL DG900 Programming Guide 4 7 ...
Page 185: ...Chapter 4 Programming Examples RIGOL DG900 Programming Guide 4 11 ...
Page 187: ...Chapter 4 Programming Examples RIGOL DG900 Programming Guide 4 13 ...
Page 199: ...Chapter 4 Programming Examples RIGOL DG900 Programming Guide 4 25 ...
Page 200: ......