Programming Guide for DM3000 Series
2-21
<range>
discrete
{0|1|2|3}
2
Explanations:
:measure:current:AC? queries the AC current. Before sending the command, please
turn on AC current measure function.
:measure:current:AC
□
<value>
sets the AC current range. At the same time the
manual mode is set.
The Table 2- 5 indicates the parameters and their meanings.
Table 2- 5 The AC current parameters and their meanings
Parameter
Range
0
20mA
1
200mA
2
2A
3
10A
MIN
20mA
MAX
10A
DEF
200mA
Example:
To set the AC current range as 10.0A:
:measure:current:AC
□
MAX or :measure:current:AC
□
3
25.
:measure:current:AC:range?
Syntax:
:measure:current:AC:range?
Function:
The command queries AC current range.
Remark:
Before sending the command, please turn on AC current measure function.
Return Format:
The query returns the range parameter: 0
、
1
、
2 or 3.(see Table 2- 5)