RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DM3000 Series
2-33
Remarks:
If the showing digit reachs 5, DEC command will be invalid. If 7,INC
invalid.
Return Format:
:measure:diode:digit? returns the current digit: 5
、
6 or 7.
Example:
To set the digit as 6:
:measure:diode:digit
□
6
51.
:measure:capacitance?
Syntax:
:measure:capacitance?
Function:
The command queries capacitance.
Return Format:
The query returns the current capacitance in the form of 8.889030e-05, and unit is F.
(for examole)
52.
:measure:capacitance
Syntax:
:measure:capacitance
□
{<
range
>|MIN|MAX|DEF}
Function:
The command sets the measure range of capacitance.
Parameters:
Name
Type
Range
Default
<
range
>
discrete
{0|1|2|3|4|5 }
2
Explanations:
The Table 2- 7 indicates the parameters and the ranges.