RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DM3000 Series
2-85
*1Note:
the scan channel includes 1~16, the channel 1~12 is connected
for measuring DCV, ACV, DCI, ACI, 2WR, FREQ, PERI, CAP, DIODE. The
channel 13~16 is connected for measuring DCV, ACV, DCI, ACI and 2WR.
*2Note:
the different measure function have the different range, thereinto:
When DCV measurement, range parameters can be 0, 1, 2, 3 or 4;
When ACV measurement, range parameters can be 0, 1, 2, 3 or 4;
When DCI measurement, range parameters can be 0, 1, 2, 3 or 4;
When ACI measurement, range parameters can be 0, 1, 2 or 3;
When 2WR measurement, range parameters can be 0, 1, 2, 3, 4, 5 or 6;
When FREQ measurement, range parameters can be 0, 1, 2, 3 or 4;
When PERI measurement, range parameters can be 0, 1, 2, 3 or 4;
When CAP measurement, range parameters can be 0, 1, 2, 3, 4 or 5.
*3Note:
ultimeter respectively shows the range as 4 ½ , 5 ½
,
6 ½ ; for 5 ½ digits
multimeter respectively shows the range as 3 ½ , 4 ½
,
5 ½ .
*4Note:
Delay Time can be integer or decimal within 0~3600, please follow the
rules of rounding if using decimal format.
3.
:scan:deletetask
Syntax:
:scan:deletetask
□
<value>
Function:
The command deletes the added task.
Explanations:
:scan:deletetask
□
<value>
deletes the added task, the value is the deleted task
number, and the range is 0~99. If the middle task is deleted, the latter task moves
forward in order.
4.
:scan:intervaltime
Syntax:
:scan:intervaltime
□
<value>
Function:
The command sets the interval time of the project cycle.
Explanations:
:scan:intervaltime
□
<value>
sets the interval time of the project cycle. The value is