Parameter
Name
Type
Range
Default
<n>
Integer
{1|2|3}
-
<numeric value>
Real
0 to the maximum current
value of the specified channel
Refer to
<def>
Discrete
{DEFault}
-
Remarks
•
If [:SOURce[<n>]] or [<n>] is omitted, the command sets the corresponding
parameter of the channel currently selected.
•
<numeric value> is the step size specified. DEFault is the default value. The
default values of <numeric value> are as shown in the table below.
Channel
Default
CH1
0.0001 A
CH2
0.0001 A
CH3
0.001 A
•
Select UP or DOWN and executes
[:SOURce[<n>]]:CURRent[:LEVel][:IMMediate]
, and the instrument will step up or down the current by the step
size set in this command.
Return Format
The query returns the step size, for example, 0.1000.
Examples
:CURR:STEP 0.1 /*Sets the step size to 0.1 A for the selected
channel.*/
:CURR:STEP? /*Queries the step size for the selected channel. The
query returns 0.1000.*/
4.9.3
[:SOURce[<n>]]:CURRent:PROTection:CLEar
Syntax
[:SOURce
[<
n>]
]:CURRent:PROTection:CLEar
Description
Clears an OCP event that occurred on the specified channel and enables the output of
the corresponding channel.
Command System