Description
Queries whether an overvoltage protection (OVP) event occurred on the specified
channel.
Parameter
Name
Type
Range
Default
<n>
Integer
{1|2|3}
-
Remarks
•
The overvoltage protection (OVP) function disables the output automatically
when the actual output voltage reaches the OVP level.
•
If [:SOURce[<n>]] or [<n>] is omitted, the command queries whether an OVP
event occurred on the current channel.
•
to query whether
an OVP event has occurred on the specified channel.
•
[:SOURce[<n>]]:VOLTage:PROTection:CLEar
that occurred on the specified channel.
Return Format
The query returns 1 or 0.
Examples
:VOLT:PROT:TRIP? /*Queries whether an OVP event occurred on the
current channel. The query returns 1.*/
4.10
:STATus Commands
:STATus commands are used to set and query the Questionable Status register and
Operation Status register.
4.10.1
:STATus:OPERation:CONDition?
Syntax
:STATus:OPERation:CONDition?
Description
Queries the value of the condition register of the Operation Status register.
Parameter
None.
Command System
68
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Summary of Contents for DP2000 Series
Page 1: ......
Page 137: ...d Exit Programming Examples Copyright RIGOL TECHNOLOGIES CO LTD All rights reserved 129...