4.12.17 :TIMEr:TEMPlet:PERIod
Syntax
:TIMEr:TEMPlet:PERIod
<
value>
:TIMEr:TEMPlet:PERIod?
Description
Sets or queries the Period of the waveform.
Parameter
Name
Type
Range
Default
<value>
Real
Refer to
Refer to
Remarks
•
<value> specifies the duration of a cycle for a waveform. All waveforms except
for ExpRise and ExpFall require a period.
•
It ranges from 1 s to 3600 s for DP932A and from 1 s to 3600 s for DP932U. After
the DP900-ARB option is installed in DP932U, it ranges from 0.1 s to 3600 s. (The
option is not available for DP932E.)
•
By default, the period of Pulse is 2 s, and the period of other waveforms is 50 s.
•
The total number of points within a period of the selected template is
determined by the period of the waveform and time interval
). The number of points within a period=period/time
interval.
Return Format
The query returns a real number ranging from 0.1 to 3600, for example, 15.0.
Examples
:TIME:TEMP:PERI 15 /*Sets the Period to 15 s.*/
:TIME:TEMP:PERI? /*Queries the Period for the selected template.
The query returns 15.0.*/
4.12.18 :TIMEr:TEMPlet:POINTs
Syntax
:TIMEr:TEMPlet:POINTs
<
value>
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
109