Description
Queries the output mode (CV, CC, or UR) for the specified channel.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CH1|CH2|CH3}
-
Remarks
•
This series power supply has three output modes:constant voltage (CV), constant
current (CC), and unregulated (UR). In CV mode, the output voltage equals to the
voltage setting value, and the output current is determined by the load; whereas
in CC mode, the output current equals to the current setting value, and the
output voltage is determined by the load. UR mode is the unregulated mode
between CV mode and CC mode.
•
If <source> is omitted, the command queries the output mode of the current
channel.
Return Format
The query returns CV, CC, or UR.
Examples
:OUTP:MODE? CH1 /*Queries the current output mode of CH1. The query
returns CV.*/
4.8.3
:OUTPut:OCP:ALAR?
Syntax
:OUTPut:OCP:ALAR?
[<
source>]
Description
Queries whether an overcurrent protection (OCP) event occurred on the specified
channel.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CH1|CH2|CH3}
-
Remarks
•
Overcurrent protection (OCP) indicates that the output is disabled automatically
when the actual output current reaches the OCP level.