4.8.14
:OUTPut:PAIR
Syntax
:OUTPut:PAIR
<
type>
:OUTPut:PAIR?
Description
Sets or queries the channel connection mode.
Parameter
Name
Type
Range
Default
<type>
Discrete
{OFF|PARallel|SERies}
OFF
Remarks
•
OFF: CH1 and CH2 are independent of each other.
•
PARallel: CH1 and CH2 are connected in parallel (internal).
•
SERies: CH1 and CH2 are connected in series (internal).
Return Format
The query returns the connection mode, for example, SERIES.
Examples
:OUTPut:PAIR SERies /*Connects CH1 and CH2 in series (internal).*/
:OUTPut:PAIR? /*Queries the connection mode. The query returns
SERIES.*/
4.8.15
:OUTPut[:STATe]
Syntax
:OUTPut[:STATe]
[<
source>,]<bool>
:OUTPut[:STATe]?
[<
source>]
Description
Sets or queries the on/off state for the specified channel.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CH1|CH2|CH3|ALL}
-
Command System
56
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.