Channel
Default
CH2
0.01 V
CH3
0.01 V
With high resolution
option
CH1
0.001 V
CH2
0.001 V
CH3
0.001 V
•
Select UP or DOWN and executes
[:SOURce[<n>]]:VOLTage[:LEVel][:IMMediate]
, and the instrument will step up or down the voltage by the step
size set in this command.
Return Format
The query returns the step size, for example, 0.100.
Examples
:VOLT:STEP 0.1 /*Sets the step size of voltage to 0.1 V for the
current channel.*/
:VOLT:STEP? /*Queries the step size of voltage for the current
channel. The query returns 0.100.*/
4.9.9
[:SOURce[<n>]]:VOLTage:PROTection:CLEar
Syntax
[:SOURce
[<
n>]
]:VOLTage:PROTection:CLEar
Description
Clears an OVP event that occurred on the specified channel and enables the output of
the corresponding channel.
Parameter
Name
Type
Range
Default
<n>
Integer
{1|2|3}
-
Remarks
•
[:SOURce[<n>]]:VOLTage:PROTection:TRIPped?
to query whether
an OVP event occurred on the specified channel.
•
Before executing the command, make sure to remove the condition that caused
the overvoltage protection on the specified channel (you can either decrease the
output voltage to below the OVP level or increase the OVP level to be greater
Command System
66
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.