RIGOL
Command Systems
Programming Guide for DS1000E, DS1000D Series
2-56
Function Explanation:
The commands set and query the vertical offset. <n> could be 1 or 2.
When Scale≥250mV, the range of <offset>is -40V~+40V;
When Scale<250mV, the range of <offset>is -2V~+2V.
Returned Format:
The query returns the setting value of <offset>.
Example:
:CHAN2:OFFS
20 Setup the vertical offset of channel 2 as 20V.
:CHAN2:OFFS?
The query returns 2.000e+01.
6.
:CHANnel<n>:PROBe
Command Format:
:CHANnel<n>:PROBe
<attn>
:CHANnel<n>:PROBe?
Function Explanation:
The commands set and query the attenuation factor of the probe. <n> could be
1 or 2; <attn> could be 1, 5, 10, 50, 100, 500 or 1000.
Returned Format:
The query returns the setting value of <attn>.
Example:
:CHAN2:PROB
10 Setup the attenuation ratio of channel 2 as 10X.
:CHAN2:PROB?
The query returns 1.000e+01.
7.
:CHANnel<n>:SCALe
Command Format:
:CHANnel<n>:SCALe
<range>
:CHANnel<n>:SCALe?
Function Explanation:
The commands set and query the vertical scale of waveform magnified by the