Output connector
The current waveform of the measured conductor is output at a constant rate.
The output connector can be connected to the BNC input connector of the
waveform measuring instrument.
NOTE
Since the output resistance is 25 Ω (RP1003C/RP1004C) or 7 Ω
(RP1005C/RP1006C), the probe must be used with a waveform
measurement instrument that has an input impedance of at least 1 MΩ.
Accurate measurement is not possible with the waveform measurement
instrument whose input impedance is 50 Ω.
If using BNC-banana plug adapters or similar to connect to input terminals
other than BNC connectors, make sure the polarity is correct.
Turn the collar until it clicks, and check that it is locked securely.
7.
Power plug
Connect this to the power supply receptacle to supply power to the sensor
terminator.