RIGOL
RP1003C/1004C/1005C/1006C User Guide
15
Location for Use
Indoor, altitude up to 2,000 m, Pollution Degree 2
Effect of External
Magnetic Fields
RP1003C: ≤20 mA (DC and 60 Hz, magnetic field of
400 A/m)
RP1004C: ≤5 mA
(DC and 60 Hz, magnetic field of 400 A/m)
Diameter of
Measurable
Conductor
5 mm
Measurable
Conductor
Insulated conductor
Guaranteed
Accuracy Period
1 year (opening/closing up to 10,000 times)
Cable Length
Sensor cable: approx. 1.5 m
Power supply cable: approx. 1 m
External Dimension
Sensor: approx. 175W×18H×40D (mm) (excluding
protrusions)
Terminator: approx. 27H×55W×18D (mm)
Mass
RP1003C: approx. 230g
RP1004C: approx. 240g
Accessories
User Guide, Probe Case
Safety
EN61010
EMC
EN61326
RP1005C
Bandwidth
DC to 10 MHz (-3 dB), refer to
Appendix 1
Amplitude-frequency Characteristics
(RP1005C)
Rise Time
≤35 ns
Maximum
Continuous Input
Range
150 Arms, refer to
Appendix 2 Relation
between Max Input Current and Frequency
(RP1005C)
Maximum Peak
Current Value
300A peak, non-continuous
500A peak, pulse width≤30μs
Gain
0.01 V/A
Amplitude
Accuracy
±1.0%rdg±1 mV, ≤150 Arms
±2.0%rdg, 150 A to 300 A peak (DC, 45 Hz to 66
Hz)
Noise
≤25 mArms (for BW 20MHz waveform measuring
instrument)
Input Impedance
Refer to
Appendix 3 Input Impedance
(Typical)
(RP1005C)
Temperature
≤±2% (during input of 55 Hz, 150 Arms within
Summary of Contents for RP1003C
Page 1: ...User Guide RP1003C 1004C 1005C 1006C Current Probe Sept 2020 RIGOL TECHNOLOGIES CO LTD ...
Page 2: ......
Page 10: ......
Page 29: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 19 RP1005C RP1006C ...
Page 31: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 21 RP1005C RP1006C ...
Page 33: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 23 RP1005C RP1006C ...