Maximum Peak
Current Value
700A peak, non-continuous
Gain
0.01 V/A
Amplitude
Accuracy
±1.0%rdg±5 mV, ≤500 Arms
±2.0%rdg, ≤700 A peak (DC, 45 Hz to 66 Hz)
Noise
≤25 mArms (for BW 20MHz waveform measuring
instrument)
Input Impedance
Refer to
Appendix 3 Input Impedance
(Typical)
(RP1006C)
Temperature
Coefficient for
Sensitivity
≤±2% (during input of 50 Hz, 500 Arms within
range of 0°C to 40°C)
Maximum Rated
Power
7.2 VA (within maximum continuous input range)
Rated Supply
Voltage
+12 V±0.5 V
Operating
Temperature and
Humidity Range
0°C to 40°C, ≤80% RH (no condensation)
Storage
Temperature and
Humidity Range
-10°C to +50°C, ≤80% RH (no condensation)
Location for Use
Indoor, altitude up to 2,000 m, Pollution Degree 2
Period of
Guaranteed
Accuracy
1 year (opening/closing up to 10,000 times)
Effect of External
Magnetic Fields
≤800 mA (in a DC or 60 Hz, 400 A/m magnetic
field)
Diameter of
Measurable
Conductor
20 mm
Measurable
Conductor
Insulated conductor
Cable Length
Sensor cable: approx. 2 m
Power supply cable: approx. 1 m
External
Dimension
Sensor: approx. 176W X 69H X 27D (mm)
Terminator: approx. 27H X 55W X 18D (mm)
Mass
Approx. 520g
Accessories
Probe Case
Safety
EN61010
EMC
EN61326
Summary of Contents for RP1003C
Page 1: ...User Guide RP1003C 1004C 1005C 1006C Current Probe Sept 2020 RIGOL TECHNOLOGIES CO LTD ...
Page 2: ......
Page 10: ......
Page 29: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 19 RP1005C RP1006C ...
Page 31: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 21 RP1005C RP1006C ...
Page 33: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 23 RP1005C RP1006C ...