14.0 TYPE-04 SPECIFICATIONS
FOR GAS PURITY
Selectable measuring gas and range
Measuring gas
(
Display indication
)
Measuring range
(Minimum digit
)
AIR CAL value
Drift amount
Spec. No.
1
Helium in AIR
(He in AIR)
0
~
100 vol%
(0.05)
0.00 vol%
0.46 vol%
SPE-1411
2
Neon in AIR
(Ne in AIR)
0
~
100 vol%
(0.1)
0.0 vol%
0.5 vol%
SPE-1442
3
Hydrogen in AIR
(H2 in AIR)
0
~
100 vol%
(0.1)
0.0 vol%
0.8 vol%
SPE-1412
4
Methane in AIR
(CH4 in AIR)
0
~
100 vol%
(0.1)
0.0 vol%
0.7 vol%
SPE-1443
5
Carbon dioxide in AIR
(CO2 in AIR)
0
~
100 vol%
(0.1)
0.0 vol%
0.8 vol%
SPE-1413
6
Ammonia in N2
(NH3 in N2)
0
~
100 vol%
(0.2)
-
6.6 vol%
1.4 vol%
SPE-1418
7
Hydrogen in N2
(H2 in N2)
0
~
100 vol%
(0.1)
3.5 vol%
0.7 vol%
SPE-1420
8
Hydrogen in CO2
(H2 in CO2)
0
~
100 vol%
(0.05)
50.20 vol%
0.38 vol%
SPE-1444
Interferometer type : Type 30MM
Indication accuracy : ±( Indication value AIR CAL. value )×0.03 ± Drift amount
AIR CAL value : Concentration indicated on the display when AIR CAL key is pressed.
Drift amount : The amount of indication drifting when temperature change ±10
℃
.
Special caution
1.
This instrument is not explosion proof. This instrument measure combustible gas around lower explosive limit
in this specification. Take enough measure for the measurement.
2.
This instrument measures extremely toxic gases. This instrument exhausts sample gas from GAS OUT. And
sample gas may leak from internal piping. Take enough measure for the measurement.