15.0 TYPE-05 SPECIFICATIONS
FOR ANAESTHETIC GASES
Selectable measuring gas and range
Measuring gas
(
Display indication
)
Measuring range
(Minimum digit
)
AIR CAL value
Drift amount
Spec. No.
1
Halothane in O2
(HALOTHANE in O2)
0
〜
6 vol% (
0.01)
1.57 vol%
0.09 vol%
SPE-1401
2
Isoflurane in O2
(ISOFLURANE in O2)
0
〜
8 vol%
(0.01)
1.65 vol%
0.09 vol%
SPE-1402
3
Sevoflurane in O2
(SEVOFLURANE in O2)
0
〜
10 vol%
(0.01)
1.70 vol%
0.09 vol%
SPE-1403
4
Desflurane in O2
(DESFLURANE in O2)
0
〜
20 vol%
(0.02)
2.10 vol%
0.12 vol%
SPE-1404
5
Halothane in AIR
(HALOTHANE in AIR)
0
〜
6 vol%
(0.01)
0.00 vol%
0.09 vol%
SPE-1397
6
Isoflurane in AIR
(ISOFLURANE in AIR)
0
〜
8 vol%
(0.01)
0.00 vol%
0.09 vol%
SPE-1398
7
Sevoflurane in AIR
(SEVOFLURANE in AIR)
0
〜
10 vol%
(0.01)
0.00 vol%
0.10 vol%
SPE-1399
8
Desflurane in AIR
(DESFLURANE in AIR)
0
〜
20 vol%
(0.02)
0.00 vol%
0.12 vol%
SPE-1400
Interferometer type : Type 30MM
Indication accuracy : ±( Indication value AIR CAL. value )×0.03 ± Drift amou
nt
AIR CAL value : Concentration indicated on the display when AIR CAL key is pressed.
Drift amount : The amount of indication drifting when temperature change ±10 .
Special caution
1. Measuring gases in this specification are not vaporized so much. The gas concentration may not reach higher
concentration in the measuring range depending of a condition of temperature and pressure.
Contact Information
http://www.gastech.com
24 Baretta Rd
Wangara Western Australia 6065
Gastech