7 Maintenance
7-1. Maintenance intervals and items
GD-D58 Series
- 38 -
<About Maintenance Services>
We provide services on regular maintenance including span adjustment, other adjustments and
maintenance.
To make the calibration gas, dedicated tools, such as a gas cylinder of the specified concentration and
gas sampling bag must be used.
Our qualified service engineers have expertise and knowledge on the dedicated tools used for services,
along with other products. To maintain the safety operation of the detector head, please use our
maintenance service.
The followings are typical maintenance services. For more information, please contact RIKEN KEIKI.
Main services
Power supply
check
: Checks the power supply voltage.
Verifies that the power lamp lights up.
Status display
check
: Checks that the status display is "run" and that the fault lamp is off.
Concentration
display check*
Verifies that the concentration display value is zero by using the zero gas.
Performs the zero adjustment if the display value is incorrect.
Flow rate check
: Checks the flow rate by using an external flow meter to check for abnormalities.
Filter check
: Checks the dust filter for dust or clogging.
Replaces a dirty or clogged dust filter.
Alarm test*
: Inspects the alarm circuit by using the alarm test function.
Checks the alarm lamps. (Checks activation.)
Checks the external alarm. (Checks the activation of the external alarm, such as a buzzer.)
Span adjustment* : Performs the span adjustment by using the calibration gas.
Gas alarm check* : Checks the gas alarm by using the calibration gas.
Checks the alarm. (Checks the alarm activation when the alarm setpoint is reached)
Checks the delay time. (Checks time to delay until the alarm is triggered.)
Checks the alarm lamps. (Checks activation.)
Checks the external alarm. (Checks the activation of external alarms, such as a buzzer and reset
signal.)
Cleaning and
repair of device
(visual diagnosis)
: Checks dust or damage on surface, cover, or internal parts of the detector head, cleans and repairs such
parts of the device.
Replaces parts which are cracked or damaged.
Device operation
check
: Uses the keys to check the operation of functions and parameters.
Replacement of
consumable parts
: Replaces consumable parts, such as a sensor, filter and pump.
* Inspect and adjust the indicator/alarm unit.