8.TREATMENT AT ABNORMAL CASE
8-1.Trouble alarm
Flashing of PW/TR light
・When PW/TR light is flashing, it is the condition that the following trouble may be caused.
Research its cause and take an appropriate action.
・The output of detector at trouble shall be as follows.
Output from GD-K71D: 0.5mA(Fix)
LED display
Trouble content
Main cause
E-00
System trouble
Memory trouble in the detector
E-01
Abnormal of sensor
connection
Connector contact failure,
Sensor cable disconnection
E-04
Abnormal function for
zero correction
Abnormality of sensor drift
E-05
Low flow
Filth of filter or deterioration
of pump
8-2.Counteraction at trouble alarm
When any trouble was found on this gas detector, contact our nearest agent or Riken Keiki.
If trouble should take place, we try best to search its cause and prevent its cause.
8-3.Before it is thought to be a trouble
It may be originated from disconnection and short circuit between units.
・Check the wirings around the related equipment as well as this detector.
There is the case that operation of detector is not correct including piping work etc.
・Re-check the detection system and contact our nearest agent or Riken Keiki.
There is the case of electrical line trouble such as shut-off of power.
・This will be treated by the re-check of stand-by battery, power line filter, insulation
transformer etc or those addition.
Summary of Contents for GD-K71D
Page 1: ...PT2E 1057 OPERATING MANUAL FOR INTELLIGENT GAS DETECTOR GD K71D...
Page 7: ......
Page 8: ......
Page 9: ......
Page 11: ......
Page 12: ......
Page 13: ......
Page 18: ......
Page 21: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 27: ......
Page 28: ......
Page 29: ......
Page 30: ......
Page 47: ......