Safety information
The Combustible Gas Monitor Model GP-1000 is a gas monitor designed to provide continuous
monitoring of exposure to combustible gas in hazardous areas.
The measuring range of GP-1000 is 0-100%LEL
The gas sample is sucked in by built-in micro pump.
Power supply is only with four series connected Alkaline Manganese AA batteries , type LR6
manufactured by TOSHIBA.
Battery exchange in a hazardous area is not performed.
Specification for safety
・
Ex ia IIC T4 Ga
・
Ambient temperature range
:
-20°C to +50°C
Electrical data
・
Powered by four series AA size alkaline batteries, type LR6 by TOSHIBA.
Certificate numbers
・
IECEx Certificate number
:
IECEx DEK 13.0090
・
ATEX Certificate number
:
DEKRA 13 ATEX 0227
List of standards
・
IEC 60079-0
:
2011
・
EN60079-0:2012+A11:2013
・
IEC 60079-11
:
2011
・
EN60079-11:2012
・
IEC 60079-26
:
2006
・
EN60079-26:2007
WARNING
・
Do not replace dry batteries in hazardous location.
・
Do not attempt to disassemble or alter the instrument.
・
Use only alkaline AA batteries, type LR6 manufactured by TOSHIBA.
・
Use only backup battery type CR1220 manufactured by Maxell.
INST. No. 0 0 0 0 0 0 0 0 0 0 0
A B C
D
E
A: Manufacturing year (0-9)
B: Manufacturing month (1-9,XYZ for Oct.-Dec.)
C: Manufacturing lot
D: Serial number
E: Code of factory
・
II 1G Ex ia IIC T4 Ga
Phone
: +81-3-3966-1113
Fax
: +81-3-3558-9110 GIII
: intdept@rikenkeiki.co.jp
Web site
: http://www.rikenkeiki.co.jp