4 How to Use
4-7. Wiring
- 23 -
4-7. Wiring
Base unit
external terminals
Indicator unit
external terminals
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(2)
(1)
(3)
(1) (2) (3)
(1) (2) (3)
(4) (5)
(6) (7)
(8) (9) (10)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(2)
(1)
(3)
(11)
(12)
Power input 100 - 120 VAC or 200 - 240 VAC (50/60 Hz)
Common alarm voltage output, 0/6/12 VDC (load current: 10 mA or less)
Common alarm contact output, no-voltage "c" contact (contact capacity: 250 VAC 2 A, resistance load)
Contact output for external buzzer (contact capacity: 250 VAC 1 A, resistance load)
Power output for external buzzer, 24 VDC (load current: 10 mA or less)
Screw size: M3
Not used
CH1 (left)
Screw size: M3
Individual alarm contact output 2
(gas alarm [standard] or fault alarm [option])
Individual alarm contact output 1 (gas alarm)
External output signal
0/6/12 VDC (load current: 10 mA)
Diffusion type
detector head
Suction type detector head
(6)
(5)
(4)
(3)
(2)
(1)