5 How to Operate
5-7. Simplified calibration (zero/span adjustment)
- 35 -
(2) Press the ZERO (
▲
) key for over 3
seconds.
After CAL is displayed, zero adjustment is
performed automatically.
When the adjustment is completed, PASS
is displayed and the span adjustment
mode is entered automatically.
* "FAIL" is displayed when zero
adjustment fails.
Press the TEST/SET key and try
again.
(3) Introduce the calibration gas.
(4) Press the TEST/SET key when the
reading is stabilized.
* SET is unavailable below 10% of the
full scale.
(5) When the reading starts blinking, adjust
the gas concentration with the
▲
/
▼
key.
(6) Press the TEST/SET key.
After CAL is displayed, gas adjustment is
performed automatically.
When the adjustment is completed, PASS
is displayed.
* "FAIL" is displayed when zero
adjustment fails.
Press the TEST/SET key and try
again.
(7) Return the MAINT switch to OFF.
CAUTION
Do not press the TEST/SET key
before the reading is stabilized.
CAUTION
Exhaust gas must be collected in
the exhaust bag or discharged
through the exhaust line.
Off
ON
Blinking
Maintenance mode active
Success
Introduce a gas and press SET when reading is
stabilized.
Adjust value with
▲
/
▼
.
Fail
Fail
Zero adjustment completed. Automatically enters
the span adjustment mode.
Success