6 Operations and Functions
6-4. Other functions
- 42 -
<Zero Follower Function>
Some types of detector heads connected to the alarm system might have sensitivity variations after being
used for a long period.
This function corrects the fluctuation of the reading at the zero point (zero drift) among the sensitivity
variations over time by a program manipulation to stabilize the zero point.
<Peak Hold Function>
In normal measurement (detection mode), the peak concentration since an alarm activation is held and
displayed on the bar.
When the full scale is exceeded, the FS bar blinks. Press BZ. STOP and then RESET to release the hold
state.
Zero Follower Off
Time
Zero point
Reading
* The zero drift of the sensor appears as the reading.
Zero Follower On
* The zero-adjustment is made periodically.
Zero point
Reading
Time