7 Maintenance
7-2. Regular maintenance mode
- 56 -
<Maintenance Mode External Output Setting 2-4. A>
* The setting is applicable only for the external output with 4-20 mA specification.
(1) Select 2-4.A from the menu screen.
(2) Press the TEST/SET key.
(3) Select 2.5/4.0/HOLD/4-20 with the
▲
/
▼
key.
(4) Press the TEST/SET key to save the setting.
External output during the maintenance
mode is set to 2.5 mA.
Last value before entering the maintenance
mode is held during the maintenance mode.
External output during the maintenance
mode is set to 4.0 mA.
External output during the maintenance
mode is set to 4 - 20 mA.