7 Maintenance
7-2. Regular maintenance mode
- 57 -
<External Output Adjustment 2-4. B>
* The setting is applicable only for the external output with 4-20 mA specification.
* The adjustment requires an ammeter separately.
(1) Connect an ammeter to the external connection terminals 8 (+) and 9 (-).
(2) Select 2-4.B from the menu screen.
(3) Press the TEST/SET key.
(4) Select 4.0/20.0 with the
▲
/
▼
key.
(5) Press the TEST/SET key.
(6) Adjust using the
▲
/
▼
key while checking the value with the ammeter.
(7) Press the TEST/SET key to save the setting.
After the setting is completed, "2-4.B" returns automatically.
External connection terminal 8
External connection terminal 7
Ammeter
Save the setting.
Adjust using the
▲
/
▼
key while "ADJ" is displayed.
Adjust the output while checking the value with the ammeter.
Save the setting.