Obtain the power consumption of GP-147 according to the following table.
(1) Power consumption of commercial power supply used for main unit (VA)
Part
Continuous
detection
A Base
unit
* Including the indicator unit (1 unit, 2 channels)
21.1 (VA)
B
Indicator/alarm unit (1 unit, 2 channels)
19.3 (VA)
C
Isobutane sensor (HW-6211)
3.8 (VA)
D
Methane sensor (HW-6239)
2.8 (VA)
E
Hydrogen sensor (NC-6244A)
1.8 (VA)
F
Suction type detector head
33.0 (VA)
(2) Power consumption while running UPS battery (W)
Part
Continuous
detection
Intermittent
detection
A Base
unit
* Including the indicator unit (1 unit, 2 channels)
2.8 (W)
2.8 (W)
B
Indicator/alarm unit (1 unit, 2 channels)
2.3 (W)
2.3 (W)
C
Isobutane sensor (HW-6211)
1.3 (W)
0.8 (W)
D
Methane sensor (HW-6239)
0.9 (W)
0.5 (W)
E
Hydrogen sensor (NC-6244A)
0.6 (W)
0.4 (W)
F
Suction type detector head
8.7 (W)
8.7 (W)
The power consumption calculating formula (P73) and data in the simplified power consumption/safety