7 Maintenance
7-3. Calibration
method
- 41 -
NOTE
If the zero adjustment failed since the zero point was significantly fluctuated from around 0.0, it
returns to 2-1 after FAIL rather than PASS is displayed. In this case, the zero adjustment has not
been completed.
<Span Adjustment "2-2">
This is used to perform the span adjustment.
CAUTION
Before starting the span adjustment, provide the detector with the calibration gas and wait until
the reading is stabilized.
WARNING
When the span adjustment is performed in the atmosphere, check the atmosphere for freshness
before beginning the adjustment. If interference gases exist, the adjustment cannot be performed
properly, thus leading to dangers when the gas leaks.