2-3 Terms of “Caution”
CAUTION
・
Do not use walkie-talkies near the wiring or by the body of this instrument. The radio
noise may affect the instruments reading. Please use radio wave emitting device
away from RI-2000R where they won’t affect the reading.
・
Please wait at least 5 seconds before restarting the instrument. If it is restarted within
5seconds from the instruments shut-down, RI-2000R might not operate properly.
・
Please check if the ball in the flow meter is above the red line before using. If the ball
is under the line, please adjust the flow rate.
・
Make sure to attach a dust filter on the gas inlet before using RI-2000R. If the filter
isn’t attached properly and dust get’s inside, it will cause false operation and damage
the instrument.
・
Please do not use this instruments output for controlling other devices. This
instrument is not a control equipment. We do not accept using RI-2000R as a control
device of any instruments.
・
There are parts that generate high temperature inside this instrument. Please do not
touch inside without care. It may cause burns and injuries.