4 Replacement of Consumable Parts
SC-8000
- 18 -
(5) Attach the sensor joint and a new sensor to the sensor cap in reverse order of sensor removal.
Insert the connector of the sensor cable to a new sensor and attach the sensor to the main unit. At this
time, be careful not to let the sensor cable be caught by the amplifier board. For the sensor with a
conversion factor label attached, tear off the product side of the label beforehand.
(6) Put the internal case cover and battery unit back in place.
For the sensor with a conversion factor label attached, peel off the conversion factor label (product
side) removed from the new sensor from release paper and then attach the label in place of the one on
the internal case cover.
(7) Turn the power ON. When "SENSOR CHANGE" is displayed, press the POWER ENTER switch to see
if the measuring mode is entered. (If the display does not appear, reattach the sensor.)
(8) Turn the power OFF and then start up in the maintenance mode again.
Perform a zero adjustment and then a span calibration.
<Replacement Procedure for Other Consumable Parts>
For the stable operation of the gas monitor and safety, a qualified service engineer is required to take care
of the replacement of the parts that requires high expertise.
Please contact RIKEN KEIKI.