9 Troubleshooting
- 49 -
SD-1DGH
9
Troubleshooting
The Troubleshooting does not explain the causes of all the malfunctions which occur on the detector. This
simply helps to find the causes of malfunctions which frequently occur. If the detector shows a symptom
which is not explained in this manual, or still has malfunctions even though remedial actions are taken,
please contact RIKEN KEIKI.
<Abnormalities on Unit>
Symptom/Display FAULT Causes
Actions
The wiring is not correct.
Connect the wiring properly.
The terminal plate is
removed.
Connect the terminal plate properly.
Abnormalities/momentary
blackout of power supply
system
Provide the rated voltage.
Take measures such as checking or adding
the UPS, power supply line filter and
insulation transformer.
The power
cannot be turned
on.
―
Cable abnormalities
(open circuit/not
connected/short circuit)
Check the wiring of detector and related
devices around it.
Abnormal
operations
―
Disturbances by sudden
surge noise, etc.
Turn off and restart the detector.
If such a symptom is observed frequently,
take appropriate measures to eliminate the
noise.
The sensor is not
connected or improperly
connected.
Check if the cable is securely fastened to the
terminal plate.
Sensor
abnormalities
E-1
z
Faults of the sensor
Replace the sensor with a new one.
The rated voltage is not
supplied to the detector.
Check the power supply, and supply the
rated voltage.
System
abnormalities
E-9
z
Abnormalities of ROM,
RAM, or EEPROM inside
of the detector
Please contact RIKEN KEIKI.