background image

10 Product Specifications

                                                                                           

10-1. List of specifications 

 

- 57 -

 

<ATEX/IECEx

Gas monitoring specifications> 

Model SD-1DOX-AS 
Type TYPE 

Detection principle 

Galvanic cell method 

Detectable gas 

O2 

Applicable gas detector head  SD-1DOX 
Gas concentration display 

LED(4digits

7segments) 

Measuring range 

0 - 5vol%/0 - 10vol%/0 - 25vol%/0 - 50vol%/0 - 100vol% 

Detection method 

Aspirator suction type 

Sampling condition 

Atmosphere equivalent (Compressed air

0.1 - 0.7MPa

Consumption Max. 

3L/min

Atmospheric conversion

Suction flow 

1.5

±

0.1L/min 

Alarm preset point 

Depend on measuring range 

Power indication 

POWER lamp lighting(green) 

Indicate accuracy 

(under an identical condition) 

Within ±0.7vol%(below 25vol% range) 
Within ±3vol%(above 25vol% range) 

Response time 

(under an identical condition) 

Within 30sec(T90)(without piping delay time) 

Gas alarm type 

Single alarm(H or L) 

Gas alarm indication 

ALM lamp lighting(red) 

Gas alarm action 

Auto-recover 

Trouble alarm

Self 

diagnosis 

System failure/Sensor failure 

Trouble alarm indication 

FAULT lamp lighting(yellow)/content display 

Trouble alarm action 

Auto-recover 

Transmission scheme 

Two-wire analog transm digital transmission(HART communication) or 
Three-wire analog transmission(in common with power supply<power 
supply,signal,common>) 

Specification of 
Transmission 

DC4 - 20mA 
(linear

load resistance less than 300

Ω

Communication scheme 

HART7 

Transmission cable 

CVVS 1.25 mm

2

 - 4-core or 3-core 

CVVS 2.0mm

2

 - 4-core or 3-core 

Transmission distance 

Less than 1.25km in case of CVVS 1.25mm

2

 

Less than 2.0km in case of CVVS 2.0mm

2

 

Power supply 

DC24V

±

10% 

Power consumption 

MAX.1.1W 

Cable connecting port 

Adapter A<NPT1/2> or adapter B<NPT3/4> or 
Pressure proof packing gland <G3/4> (Compatible cables 

φ

9.6 - 13.0mm in 

outer diameter) 

Tube connecting port

*1

 

Gas inlet

Rc1/4(with PP union<for PTFE pipe> for O.D

φ

6-1

Gas outlet

Bs union(for Cu pipe) for O.D

φ

8-1

 

Compressed air inlet

BS union(for Cu pipe) for O.D

φ

8-1

 

Initial clear 

Approx.25sec 

Operating temperature 

-10 - +40

(non-rapidly-vary) 

Operating humidity 

Less than 95%RH(non-condensing) 

Structure 

Wall mounting sampling unit 

Explosion-proof structure 

Flame proof structure 

Explosion-proof grade 

2G Ex db 

C T6 Gb(ATEX) / Ex db 

C T6 Gb(IECEx) 

Outer dimension 

Approx.400(W)×265(H)×102(D)mm(projection excluding) 

Weight Approx.6.3kg 
Color Munsell 

7.5BG5/2 

*1 When a half-union for gas IN side is changed, check also the external filter (IN/OUT diameter) carefully. 

 
 

 

Summary of Contents for SD-1DOX-AS

Page 1: ...PT2E 2625 Smart Transmitter Gas Detector Head SD 1DOX AS TYPE H Operating Manual PT2E 179 PT2 179...

Page 2: ...addition it may be fluctuated by environmental temperature humidity etc changes in the installation site 2 The alarm must be set within a range where the performance of the detector can be ensured The...

Page 3: ...o wire 21 4 7 How to tube 25 5 How to Operate 26 5 1 Preparation for start up 26 5 2 Basic operating procedures 26 5 3 How to start the gas detector 27 5 4 Modes 28 5 5 User mode 29 5 6 Flow Rate Adju...

Page 4: ...s concentration is over the setting value The detector is a safety unit not an analyzer or densitometer which performs quantitative qualitative analysis measurement for gases Please fully understand t...

Page 5: ......

Page 6: ......

Page 7: ......

Page 8: ......

Page 9: ......

Page 10: ......

Page 11: ...standard accessories Main Unit Standard Accessories Operating manual one Dedicated handling lever one Dedicated control key The supplied quantity depends on the number of units to be delivered 1 to 10...

Page 12: ......

Page 13: ......

Page 14: ...o switch menus or change a value DOWN 5 Power lamp Power lamp Detection mode It lights up in green Maintenance mode It blinks in green 6 Alarm lamp Alarm lamp It lights up in red when the alarm setpoi...

Page 15: ......

Page 16: ......

Page 17: ......

Page 18: ......

Page 19: ...ning is impossible the following protective measures can be taken Protection against lightning Take appropriate measures in accordance with the importance of the facilities and the environment Provide...

Page 20: ......

Page 21: ......

Page 22: ......

Page 23: ......

Page 24: ......

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ......

Page 29: ......

Page 30: ......

Page 31: ......

Page 32: ......

Page 33: ......

Page 34: ......

Page 35: ...d if not needed Display Operation Oxygen Concentration Display In case of over the detection range Over Scale is displayed on the LED Power Indicator Lamp POWER Green During operation this lights up c...

Page 36: ...cess the gas leak point equipped with a protective gear to avoid dangers caused by possibly remaining gases and check whether gases remain or not by using a portable gas detector Check that the point...

Page 37: ......

Page 38: ......

Page 39: ...as a safety unit Maintenance item Maintenance content Daily mainte nance Monthly mainte nance Regular mainte nance Power Supply Check Check that the power lamp lights up Concentration Display Check C...

Page 40: ...r to find abnormalities Checks the flow rate by using an external flow meter to verify the correctness of the flow rate indicator on the gas detector If the flow rate is incorrect performs the flow ra...

Page 41: ......

Page 42: ......

Page 43: ......

Page 44: ......

Page 45: ......

Page 46: ......

Page 47: ......

Page 48: ......

Page 49: ......

Page 50: ......

Page 51: ......

Page 52: ......

Page 53: ......

Page 54: ......

Page 55: ...vide the rated voltage Take measures such as checking or adding the UPS power supply line filter and insulation transformer Cable abnormalities open circuit not connected short circuit Check the wirin...

Page 56: ...ently take appropriate measures to eliminate the noise Sudden change in the environment When the environment temperature etc changes suddenly the detector cannot adjust to it and is affected by it In...

Page 57: ...ition Less than 1vol against alarm preset point Alarm delay time 1 under an identical condition By anoxia alarm Alarm preset point 18vol Less than 5sec When introducing 10 11vol gas Without piping del...

Page 58: ...let Bs union for Cu pipe for O D 8 1 Compressed air inlet BS union for Cu pipe for O D 8 1 Initial clear Approx 25sec Operating temperature 10 40 non rapidly vary Operating humidity Less than 95 RH no...

Page 59: ...analog transmission digital transmission HART communication or Three wire analog transmission in common with power supply power supply signal common Specification of Transmission DC4 20mA linear load...

Page 60: ...o wire analog transmission digital transmission HART communication or Three wire analog transmission in common with power supply power supply signal common Specification of Transmission DC4 20mA linea...

Page 61: ...10 Product Specifications 10 1 List of specifications 59 Outline Drawings Adapter A NPT1 2...

Page 62: ...1a Non exciting at normal exciting at alarm or exciting at normal non exciting at alarm Contact capacity AC250V 0 5A DC30V 0 5A resistive load Transmission scheme Two wire analog transmission digital...

Page 63: ...ver Transmission scheme Two wire analog transmission digital transmission HART communication or Three wire analog transmission in common with power supply power supply signal common Specification of T...

Page 64: ...o recover Transmission scheme Two wire analog transmission digital transmission HART communication or Three wire analog transmission in common with power supply power supply signal common Specificatio...

Page 65: ...10 Product Specifications 10 1 List of specifications 63 Outline Drawings Cable gland G3 4...

Page 66: ......

Page 67: ...power This is a function to prevent triggering alarm during that time Full scale Maximum value of the detection range Calibration Adjusts the readings to the calibration gas concentration value by usi...

Page 68: ...0 First issue 3 22 2018 1 Declaration of Conformity 12 23 2019 2 Declaration of Conformity 4 1 2020 3 Safety information Declaration of Conformity 8 1 2020 4 Safety information Declaration of Conformi...

Page 69: ......

Reviews: