Parameter
Description
Range
0
Settings to default
0 = User database
1 = Runtime database
2 = Calibration
3 = Ethernet
4 = Licensing database
Example
Command
Answer
Description
S01;
Select unit 1
DFT1;
0 CRLF
Reset the user database
21.3.16
DPF, DPS: Passcodes
The Safe Passcode protects against misuse by the operator, and prevents access to Safe Setup without entering
the correct passcode. It does not block changes to settings from the serial ports. The Full Setup can be
accessed normally.
The Full Passcode protects against the access to the Full setup through the keys and blocks the change of all
trade relevant items through the serial ports. To prevent attempts to crack the full passcode via the serial
ports, the unit will only permit 5 unsuccessful attempts. After this any further attempts to enter the full
passcode will be ignored until the unit has been restarted.
Sending the passcode in the free stage defines the passcode and locks the unit, sending the passcode in locked
stage opens the unit again until it is deselected.
General
No. of parameters
1
Save changes
With
TDD1
Increment Trade Counter
No
Parameter Details
Parameter
Description
Range
Default
1
Passcode
000000 - 999999
000000
Note
0 is not a valid passcode, and should instead be written to the parameter to clear
the existing passcode.
Example
131
C500-600-2.0.3
Summary of Contents for C500 series
Page 1: ...C500 Series Digital Indicator Reference Manual ...
Page 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Page 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Page 123: ...118 C500 600 2 0 3 ...
Page 182: ...177 C500 600 2 0 3 ...